WeissTechnikHigh Power Rapid Temperature Change Chamber

ClimeEvent

ClimeEvent is a high-performance, high-power rapid temperature change tester that meets a wide range of the world's major test standards. Equipped with the industry's highest level of rapid temperature change performance of up to 25°C/min, it precisely evaluates the reliability of products and components by subjecting them to severe temperature stress. It uses R449A refrigerant, which has a global warming potential (GWP) about one-third lower than conventional refrigerants, and complies with the latest F-gas regulations. With a wide variety of models ranging in capacity from 270 to 1300 liters, it is ideal for a wide range of applications from R&D to mass production verification.

WeissTechnikindicates word or phrase being defined

Weiss Technik

Basic performance

  • Temperature change rateRapid temperature change of up to 25°C/min (compliant with IEC 60068-3-5)
  • Temperature range:Wide temperature range: -72°C to +180°C.
  • Frost-free operation10°C/10%RH to 95°C/95%RH
  • Allowable heat generation loadUp to 8 kW
  • Equipped with greennmode®.Power consumption reduced by up to 40%
  • Low GWP refrigerantAdopted: R-449A
  • user-friendly mechanismJapanese display control panel. Highly visible large observation, etc.
  • maintainabilityLow wick replacement frequency due to unique water supply mechanism
  • test capacityFrom a wide range of capacities from 270L to 1300L
Product Catalog
specification sheet

Product Features

Rapid temperature change by high power specification 25°C/min

Rapid temperature change in the range of 10°C to 25°C/min. High performance in all specifications, including a wide range of test conditions and short test time, and meets European and other global test standards such as IEC 60068-3-5 with ease.

Frost-free over a wide temperature and humidity range

A unique humidity control system enables frost-free operation over a wide temperature and humidity range for extended periods of time. The temperature and humidity range can be extended to 10°C / 10%RH by using the optional compressed air introduction mechanism.

Good humidity tracking over a wide temperature range

Rapid temperature change conditions while maintaining stable humidity. All ClimeEvent series have high basic performance that meets test standards that require advanced control, such as IEC60068-2-30.

Wide temperature range up to 8 kW allowable heating load

The temperature range is from -20°C to +100°C and the allowable heat generation load is up to 8 kW.

operability

  • Comfortable testing environment and high performanceHighly controlled air guidance system, designed to prevent water droplets from adhering to the sample.
  • Highly sealed, highly corrosion-resistantThe following is a list of the most important features of the product:: Stainless steel sheet and special selective sealing for high airtightness and high corrosion resistance by choice.
  • Simple wick exchange systemWick: Wick is laid out in the lower front part of the tank. The front door can be opened for easy access for easy replacement. In addition, the unique water supply mechanism reduces the frequency of replacement.
  • Large observation windowThe largest area in its class (W420 x H570mm). Unique four-tank glass structure for outstanding visibility.
  • Touch panel "WEBSeason®"Color TFT touch display with Japanese display and intuitive interface. Intuitive operation for both advanced program settings and data analysis. Multi-user functionality allows assignment of user privileges and simultaneous access, as well as remote access from PCs, smartphones, and tablets. Test results can be downloaded in CSV format.
  • S!MPATI® (optional)Data management software. Test equipment and measurement results can be documented and edited together in an easy-to-understand manner. Up to 99 test instruments can be managed.

Environmentally Friendly

  • Energy-saving performance: greennmode® is installed as standard equipment for efficient operation of heating and cooling systems. Reduced power consumption by up to 40% and lower testing costs.
  • Low GWP refrigerant: New refrigerant R-469A is used, with a GWP value of 1357, which is more than 90% less than R-23 and 75% less than R-508A, significantly reducing the burden on the global environment.
  • Regulatory compliance: Meets F-gas regulations and can meet future environmental standards.

specification sheet

A wide range of variations are available to suit your sample size and test purpose.

How to read the model

c

Temperature/Humidity

Temperature and humidity:c

Temperature:t

/

1300

Internal dimensions (WxHxD mm)

1300 1100×950×1325

800 1100×950×800

480: 800×950×650

270580×750×615

/

70

Temperature Type

70: -72°C to +180°C

40: -42°C to +180°C

/

25

Temperature change rate

2525°C / min.

2020°C / min

1515°C / min

1010℃ /min

View Specifications (PDF)

(data) item

C/270/70/15

C/480/70/15

C/800/70/15

C/1300/70/15

C/270/70/25

C/480/70/25

C/800/70/25

C/1300/70/25

Temperature range:

-72°C to +180°C

Humidity range

(Settable temperature range)

10%~98%

(10°C to 95°C)

Temperature rise time

Approx. 19 min.

-72°C to +180°C

Approx. 16 min.

-72°C to +180°C

Approx. 14 min.

-72°C to +180°C

Approx. 12 min.

-72°C to +180°C

Approx. 11 min.

-72°C to +180°C

Approx. 12 min.

-72°C to +180°C

Approx. 10 min.

-72°C to +180°C

Temperature fall time

Approx. 26 min.

+180°C to -70°C

Approx. 31 min.

+180°C to -70°C

Approx. 19 min.

+180°C to -70°C

Approx. 26 min.

+180°C to -70°C

Approx. 15 min.

+180°C to -70°C

Approx. 16 min.

+180°C to -70°C

Temperature change rate (rise)

15°C/min

-45°C to +155°C

17°C/min

-45°C to +155°C

16°C/min

-45°C to +155°C

25°C/min

-45°C to +155°C

Temperature change rate (descending)

18°C/min

+155°C to -45°C

15°C/min

+155°C to -45°C

15.5°C/min

+155°C to -45°C

14.5°C/min

+155°C to -45°C

25°C/min

+155°C to -45°C

23°C / min

+155°C to -45°C

Temperature and humidity fluctuations

Temperature operation ±0.1K to ±0.5K|Temperature/Humidity operation ±0.1K to ±0.3K/1 to 3%RH

Temperature and humidity gradient

During temperature operation ≦4°C

During temperature operation ≦2°C

Space temperature and humidity deviation

Temperature operation: ±0.5°C to ±2.0°C

Allowable heat generation load

(during temperature operation)

8000W

(SV -20°C to 100°C)

Allowable heat generation load

(in humidity operation)

400W

(SV 25°C to 95°C,),

~(~90%RH)

500W

(SV 25°C to 95°C, ~90%RH)

400W

(SV 25°C to 95°C,),

~(~90%RH)

500W

(SV 25°C to 95°C, ~90%RH)

Operable ambient temperature range

+10°C to +35°C

Internal dimensions (W x H x D mm)

580 x 750 x 615

800 x 950 x 650

1100 x 950 x 800

1100 x 950 x 1325

580 x 750 x 615

800 x 950 x 650

1100 x 950 x 800

1100 x 950 x 1325

Content volume (liter)

267

494

836

1384

267

494

836

1384

External dimensions (W x H x D mm)

895 x 1950 x 2155

1115 x 2090 x 2680

1415 x 2085 x 2885

1415 x 2085 x 3410

895 x 1950 x 2155

1115 x 2535 x 2985

1415 x 2085 x 3285

1415 x 2085 x 3810

Weight (kg)

Approximately 835

Approx. 1350

Approximately 1700

Approximately 1950

Approx. 980

Approx. 1600

Approx. 2050

Approx. 2300

Exterior Steel Plates

Electro-galvanized steel sheet

(Powder coating finish)

Interior Steel Plates

Stainless steel plate

(SUS304+SUS316)

Maximum number of shelves

8

11

8

11

ventilator

turbofan

refrigerator

Water-cooled 2-way refrigeration system

Refrigerant used

R449A+R23

Heating Heater

sheathed heater

humidifier

sheathed heater

cooler

Plate Fin Coil

Control Sensors

Pt100Ω

Control method

PID control

Setting/Indicator

Door Attached

Display method

7-inch LCD

Setting method

Interactive method with LCD touch panel

memory capacity

1GB

Ethernet

1 pc (100/1000MB)

USB

1 piece

Power supply (with transformer)

AC200V 3φ 50/60Hz Voltage fluctuation ±10%

humidified water

Ion-exchanged water or distilled water (5 to 20 μS/cm)

Water temperature 18°C, inlet/outlet temperature difference 10 K (L/min)

33

52

75

47

108

Water temperature 28°C, inlet/outlet temperature difference 5K (L/min)

65

105

150

93

217

Cooling water pressure

0.25 to 0.6 MPa Outlet/outlet differential pressure: 0.2 MPa or higher

Cooling water temperature

+12°C to +28°C

Cooling water connection Inlet/outlet pipe diameter

Inlet 1”|Outlet 1” (Female thread)

Inlet 1 1/2”|Outlet 1 1/2” (female thread)

Maximum load current with transformer (A)

62

86

128

84

148

174

Fuse current in equipment (A)

63 (400V AC)

80 (400V AC)

63 (400V AC)

100 (400 VAC)

Maximum heat removal (kW)

2

3.5

2

3.5

safeguard

Fan motor overload protector, Empty burnout protector, Over-temperature/over-cooling protector for sample protection, Over-temperature protector for equipment protection, Interlock terminal for sample energization, Refrigeration machine pressure switch, Refrigeration machine overheat protector, Power supply reverse phase protector, Water level drop alarm for water supply tank

body (of a machine)

RJ45 connector for Ethernet, USB connector, digital I/O (4 points each, dedicated connector), over-temperature protector for sample protection, observation window (effective dimensions: W420 x H570), round measurement holes (right side: φ125, left side: φ50 each), etc.

accessory

Instruction manual (1 set), water supply tank (20L), shelf plate (1 piece), silicon plug for round measurement hole (for φ50 and φ125, 1 piece each), spare wick gauze (2m), inspection report (1 set)

Supported options

Water purifier, cable holes, plugs for cable holes, floor load capacity, shelf netting for various load capacities, stainless steel coils, low silicon specifications, test area lighting, additional safety devices, emergency stop switch, signal tower (two colors), dehumidification mechanism to expand controllable temperature and humidity range, humidity sensor specifications, sample temperature control, green mode, air-cooled version, spare parts package, etc.