Basic performance
- Refrigeration equipment: air-cooled and water-cooled types are lined up.
- Internal dimensions (W x H x D mm): 370 x 330 x 400, 370 x 500 x 400, 650 x 500 x 400, 700 x 500 x 600
- Contents: 48L, 74L, 130L, 210L
- Low temperature test temperature range: -65°C to 0°C (-80°C to 0°C for water-cooled type)
- High temperature test temperature range: +60°C to +200°C
- Pre-cooling temperature range: -75°C to 0°C (-80°C to 0°C for some water-cooled types)
- Preheating temperature range: +60°C to +225°C
operability
Large LED and color LCD touch panel
Large LEDs on the top of the ETACOM allow operation and abnormal conditions to be checked at a glance, and the 7.5-inch wide color LCD touch panel provides intuitive and easy-to-understand operability.
Automatic storage of test chamber temperature
Automatically saves test room temperatures even when not in operation. The history can be checked and downloaded on a monitor or USB memory stick.
Various display functions
Graphical display of program progress and test conditions facilitates confirmation.
Configuration functions that support high performance
Temperature, exposure time, and number of cycles can be easily entered and edited. Elapsed cycles can be managed individually with a counter with reset.
Maintenance Management Functions
Sensors placed around the chiller constantly monitor the operational status. In the event of trouble, the operating conditions are instantly saved, enabling quick identification of the cause.
Optional additional features
Flexible test operation is supported by functions such as the ability to change exposure time and the ability to interrupt specified cycles.
Main functions (list)
- Number of program memory: Max. 30 types (No.1 to No.30)
- Upper and lower limit temperature alarm function
- weight function
- Auto start function (operation/standby)
- End time display (calendar display of year, month, date, hour and minute)
- Test room temperature data download function
- maintenance management monitoring function
- High/low temperature soaking time: up to 120 min.
- Room temperature soaking time: up to 99 minutes
- Alarm history storage function
- Troubleshooting display screen
- Time signal output (250 VAC 3A/3 channels)
- Interlock contact (AC220V 10A)
- Test end output (AC250V 3A)
- External alarm output (AC250V 3A)
- Communication function (RS485: optional)
Environmentally Friendly
Low GWP refrigerant R-448A adopted
The use of refrigerant R-448A (GWP value 1387) reduces environmental impact by approximately 65% compared to conventional R-404A. It also meets the 2025 target value (1,500 or less) under the Fluorocarbons Emission Control Law, and is environmentally friendly and safe.
air-tank temperature cycle tester
WINTECH NEO (300°C specification)
The WINTECH series with 300°C specifications is now available for reliability evaluation of power devices for automotive and other industrial applications with high voltage and high current. The WINTECH series has improved heat resistance and insulation performance during high-temperature operation, achieving both basic performance and safety at a high level.
Basic performance
- Chiller: air-cooled (NT3702A), water-cooled (NT3702W)
- Internal dimensions (W x H x D mm): 370 x 450 x 400
- Contents: 66L
- Low temperature test temperature range: -65°C to -10°C
- High temperature test temperature range: +80°C to +300°C
- Pre-cooling temperature range: -75°C to -20°C
- Preheating temperature range: +90°C to +350°C
- Packing for high temperature damper
Low-temperature recovery time shortening function
When shifting from "high temperature rinsing" to "low temperature rinsing" during a 2-zone test, outside air is introduced until the set temperature is reached, significantly shortening the recovery time. This is particularly effective in accelerating the time required for specimens to return to temperature under conditions with a large temperature range or under load.
Safety Considerations
Insulation is applied to areas that are easily touched by human hands, such as cable measurement holes that are prone to high temperatures, to ensure safety even if accidentally touched. Safety is ensured down to the smallest detail.
Liquid bath thermal shock tester
WINTECH NEO LIQUID Series
This liquid bath type thermal shock tester is effective in reducing test time for evaluation of mounted products. This tester is best suited for evaluating the reliability of connections of mounted automotive electronics products, and its testing method is designed to meet the needs for shortening the time required for the development of new vehicles. The inside of the chamber and the specimen transfer space are welded together to ensure a high level of confidentiality. Furthermore, a vapor recovery device is provided as standard equipment to significantly reduce the consumption of brine solution.
Air cylinder method
To protect the specimens during the test, an air cylinder system is used to drive the specimen basket. The specimens are safely protected even in the event of power failure or malfunction, and the door cannot be opened or closed during the test to ensure safety.
increased load-bearing capacity
Sample baskets are available in 3 kg (LT20, 60) and 6 kg (LT80) load capacities, and a larger basket for A4-size samples is also available for a variety of sample evaluations.
Brine solution automatic feeder
An automatic brine supply function is provided as standard equipment to enhance test stability. Both low and high temperature tanks maintain the liquid level and automatically prepare for the next test.
Removable sample basket
A removable sample basket is used to improve workability. Samples can be smoothly replaced.
Brine solution cleaning system (optional)
Brine solution is kept clean even during long-term tests or tests with frequent sample changes. Built-in cleaning system for quick and easy cleaning.
System Upgrade (optional)
System linkage between thermal shock testing and continuity reliability evaluation tester (MLR22) is possible. Connection reliability testing can be performed efficiently.
option
It can be customized with a variety of options that improve functionality and ease of use.
Sample temperature control
Even if the material or mass of the sample changes, the exposure time can be accurately controlled.
paperless recorder
Data can be saved on an SD card and imported to a PC using dedicated software.
RS485 interface
Remote operation and data management are possible through connection to a PC.
Rectangular Cable Hole
One or two additional locations can be added on both sides.
Power Quantity Monitor
The amount of power consumption and CO2 emissions of the testing equipment can be managed.
Air compressors
Built-in dedicated box to drive damper and door opening/closing mechanism.
Round measurement hole
Equipped with one location on the left side as standard, additional locations can be added.
Sample baskets and receptacles
Mesh sample baskets and receiving fittings are available.
Downwind Sensor
Additional sensors for a wide range of testing conditions.
Cycle counter (with separate instrument)
An additive counter that can retain the number of cycles even when the power is turned off.
AC100V outlet
Can be used as a power supply for measuring instruments (rated 3A).
Function to change exposure time
It is possible to change the exposure time during the test.
Designated cycle interruption function
The test can be automatically interrupted at each set cycle and resumed as needed.
specification sheet
Temperature Cycling Tester (Gas Tank Type)
WINTECH NEO AIR
model (e.g. of a vehicle) | NT551A | NT1051A | NT1251A | NT1051W | NT1251W | NT2051W | NT3702A | NT3702W |
Temperature cycle method | Cold/heat air switching system with damper (sample stationary system) | |||||||
Applicable Tests | 2-zone and 3-zone temperature cycling tests | |||||||
Low temperature test temperature range | -65°C to 0°C | |||||||
High temperature test temperature range | +60°C to +200°C | +80°C to +300°C | ||||||
Return Time to Temperature | Within 5 minutes | Within 4 minutes | Within 10 minutes (SR function ON) | Within 7 minutes (SR function ON) | ||||
Low temperature/time | -40°C/30 min. | -55°C/30 min. | -65°C/30 min. | -40°C/30 min. | ||||
Room temperature soaking time | – | 5 min. | 0 minutes | |||||
High temperature exposure temperature / time | 125°C/30 min | 150°C/30 min | 250°C/30 min | 300°C/30 min | ||||
sample | Plastic mold IC 2.5 kg + feed basket, receiving fixture 1 kg | Plastic mold IC 2.5 kg + feed basket, receiving fixture 2 kg | Plastic mold IC 2.5kg + feed basket, receiving fixture 2.8kg | Plastic mold IC 2.5 kg + sample basket, receiving fixture 1 kg | Plastic mold IC 5kg + sample basket, receiving bracket 2kg | Plastic mold IC 5kg + sample basket, receiving fixture 2.8kg | Aluminum piece 1.25kg | Aluminum strip 2.5kg |
Control sensor position | upwind | |||||||
Pre-cooling temperature range | -75°C to 0°C | ₋80°C to 0°C | -75°C to -20°C | |||||
Preheating temperature range | +60°C to +225°C | +90°C to +350°C | ||||||
Guaranteed ambient temperature | 23°C±5°C | |||||||
Ambient operating temperature | +10°C to +35°C | +10°C to +32°C | +5°C to +35°C | +10°C to +35°C | +5°C to +35°C | |||
Tank Structure | Three-tank type (test chamber, high-temperature chamber, low-temperature chamber) | |||||||
Test chamber dimensions (W x H x D mm) | 370 x 330 x 400 | 370 x 500 x 400 | 650 x 500 x 400 | 370 x 500 x 400 | 650 x 500 x 400 | 700 x 500 x 600 | 370×450×400 | |
External dimensions (W x H x D mm) Not including protrusions | 1150 x 1800 x 1325 | 1150 x 1940 x 1770 | 1430 x 1940 x 1720 | 1150 x 1940 x 1620 | 1430 x 1940 x 1470 | 1480 x 1940 x 1670 | 1150 x 1980 x 1720 | 1150×1980×1570 |
Main unit weight (kg) | Approx. 730 | Approx. 1050 | approximately 1230 | Approx. 1020 | Approx. 1180 | Approx. 1280 | Approx. 1150 | Approx. 1000 |
Test chamber load capacity (kg) | 30 (Equal distribution load) | 30 (Equal distribution load) | 100 (Equal distribution load) | 30 (Equal distribution load) | 100 (Equal distribution load) | 100 (Equal distribution load) | 30 (Equal distribution load) | 30 (Equal distribution load) |
Allowable number of sample baskets | six flat objects (e.g. tickets, pieces of cloth, pieces of cloth, etc.) | seven (used in legal documents) | six flat objects (e.g. tickets, pieces of cloth, pieces of cloth, etc.) | |||||
Allowable load of material basket | 5.0kg/sheet | 7.5kg/sheet | 5.0kg/sheet | |||||
door opener | single swing door | |||||||
measurement hole | φ50mm round measurement hole (1 pcs. on the left side of the main unit) | 30mm x 100mm square measurement hole (1 pcs. on the left side of the main unit) | ||||||
Control method | PID control method (upwind) | PID control method | ||||||
power-saving method | ECO operation (pre-cooling/pre-heating suspension method with learning function) | |||||||
Setting method | LCD touch panel system | |||||||
Display method | 7.5″ color LCD display | |||||||
Display resolution | 1°C | |||||||
Number of program registrations | Up to 30 types (Program No. 1 to No. 30) | |||||||
Temperature Cycle | Max. 9999 cycles | |||||||
Exposure time setting | High/low temperature rinsing max. 120 min|Rinsing at room temperature max. 99 min | |||||||
memory capacity | 16GB | |||||||
interface | USB2.0 TYPE-A connector | |||||||
Ancillary Functions | Auto start (operation, standby), automatic defrosting, wait, self-monitoring monitor, test end time display, elapsed cycle display, trouble handling display, automatic pre-cooling and preheating temperature setting, alarm history storage, test end status selection (cycle stop, stop after defrost, continue test preparation after cycle end), cycle with reset counter (4 conditions), end-of-test output, time signal output (3 points), upper and lower limit temperature alarm function, trouble message display, graphic display, USB port, pause, cold chamber drying function, etc. | Auto start (operation, standby), automatic defrosting, wait, self-monitoring monitor, test end time display, elapsed cycle display, trouble handling display, automatic pre-cooling and preheating temperature setting, alarm history storage, test end status selection (cycle stop, stop after defrost, continue test preparation after cycle end), cycle with reset Counter (4 conditions), end-of-test output, time signal output (3 points), upper and lower limit temperature alarm function, trouble message display, graphic display, USB port, pause, SR function, cold room drying function, etc. | ||||||
Upwind and downwind temperature sensors | Pt100 (JIS C1604) | – | ||||||
Temperature sensor | – | Pt100 (JIS C1604) | ||||||
Refrigeration method | dual refrigeration system | |||||||
refrigerator | Air-cooled sealed reciprocating compressors | Air-cooled Hermetic Scroll Compressors | Scroll type compressor | |||||
refrigerant | R-448A+HFC23 | R-404A+R-23 | ||||||
heater | Strip Wire Heaters | |||||||
ventilator | sirocco fan | |||||||
Damper drive unit | air cylinder | |||||||
safeguard | Earth leakage circuit breaker, blower thermal relay, power supply reverse phase prevention relay, high temperature room temperature overload preventer, test room temperature overload preventer, low temperature room temperature overload preventer, test room temperature upper/lower limit alarm device, refrigerator 1 and refrigerator 3 overload relay, refrigerator 1 and refrigerator 3 high/low pressure switch, proximity switch for damper, door open detection mechanism Sensor disconnection detection function, ambient temperature abnormality detection function, external alarm output, interlock terminal | Earth leakage circuit breaker, blower thermal relay, power supply reverse phase prevention relay, high temperature room temperature overload preventer, test room temperature overload preventer, low temperature room temperature overload preventer, test room temperature upper/lower limit alarm device, refrigerator 1 and refrigerator 3 overload relay, refrigerator 1 and refrigerator 3 high/low pressure switch, proximity switch for damper, door open detection mechanism Sensor disconnection detection function, cooling water temperature abnormality detection function, cooling water breakage detection function, external alarm output, interlock terminal | Earth leakage circuit breaker, blower thermal relay, blower reversal prevention relay, high temperature chamber temperature overload preventer, test chamber temperature overload preventer, low temperature chamber temperature overload preventer, test chamber temperature upper/lower limit alarm device, refrigerator 1 and refrigerator 3 overload relay, refrigerator 1 and refrigerator 3 high/low pressure switch, proximity switch for damper, door lock mechanism, sensor disconnection detection function, ambient temperature abnormality detection function, external alarm output, interlock terminal | |||||
power (button on TV, etc.) | AC200V 3 phase 50/60Hz Power supply fluctuation ±10 | |||||||
Maximum power consumption (kVA) | 19 | 30 | 32 | 35 | 44 | 48 | 35 | 40 |
Maximum load current (A) | 55 | 87 | 93 | 100 | 125 | 138 | 101 | 113 |
Operating current (A) | 26 | 49 | 50 | 56 | 63 | 65 | 71 | 80 |
Refrigeration machine exhaust heat (kW) | 7 | 11 | ₋ | 11 | – | |||
Cooling water volume (liter/min) | ₋ | 38 (water temperature 25°C)/55 (water temperature 32°C) connection bore 1’1/4B | 47 (water temperature 25°C)/82 (water temperature 32°C) connection bore 1’1/4B | – | ||||
Coolant inlet pressure (MPa) | ₋ | 0.2 to 0.5 | – | |||||
circulating volume | – | 55 liter/min (water temperature 32°C) | ||||||
Water supply water pressure | – | 0.2 to 0.5 MPa | ||||||
air | 0.4 to 0.7MPa (dry air) connected to the back of the main unit with a φ6mm (outer diameter) air hose | |||||||
drainage (e.g., drainage during defrosting) | Atmospheric release (1 location) φ8mm (I.D.) x φ11mm silicone hose | |||||||
accessory | Silicone stopper for measurement hole (2 pcs), instruction manual (1 copy), warranty card (1 copy), inspection certificate (1 copy) | Silicone plugs for measuring holes (2), Y strainer for cooling piping (1), instruction manual (1 copy), warranty card (1 copy), inspection certificate (1 copy) | Silicone plug for measurement hole (1 pc), heat-resistant plug for measurement hole (1 pc), instruction manual (1 copy), warranty card (1 copy), inspection certificate (1 copy) | |||||
Thermal shock tester (liquid tank type)
WINTECH NEO LIQUID
model (e.g. of a vehicle) | LT20W | LT60W | LT80W |
system | Sample basket moving type (2- or 3-zone switchable) | ||
Low temperature test temperature range | -65°C to 0°C | ||
High temperature test temperature range | +50 to +200°C | ||
Temperature fluctuation range | ±0.5°C | ||
Pre-cooling temperature range | MAX -70°C | ||
Pre-cooling temperature fall time | Within 60 minutes +20°C to -65°C | Within 90 minutes +20°C to -65°C | |
Preheating temperature range | MAX +200°C | ||
Preheat temperature rise time | Within 60 minutes From +20°C to +150°C | Within 90 minutes +20°C to -150°C | |
Ambient operating temperature | +5°C to +40°C | ||
Sample | Plastic Mold IC 0.5kg | Printed circuit board 1.5kg | Printed circuit board 2.0kg |
Temperature (low-temp/high-temp tank) | Low temperature tank -65°C|High temperature tank +150°C | ||
exposure time | 5 min. | ||
Return Time to Temperature | 2 minutes | 3 minutes | |
Temperature measurement position | Near the sample basket | ||
Tank Structure | 3-tank type (low temperature, high temperature, room temperature) | ||
Sample basket dimensions (W x H x D mm) | 120 x 120 x 150 | 170 x 170 x 250 | 220 x 220 x 310 |
External dimensions ((W x H x D mm)) | 1480 x 1880 x 1300 | 1740 x 1880 x 1500 | 1810 x 1850 x 1500 |
Main unit weight (kg) (excluding brine solution) | Approx. 1000 | approximately 1500 | Approximately 1700 |
Fluorine inert liquid volume (high/low temperature side) | High temperature side Approx. 33 liters Low temperature side Approx. 42 liters | High temperature side Approx. 48 liters Low temperature side Approx. 62 liters | High temperature side Approx. 55 liters Low temperature side Approx. 75 liters |
Allowable load of sample basket (kg) | 2 | 6 | |
Control method | PID control method | ||
Setting method | LCD touch panel interaction | ||
Display method | LCD graphic display | ||
Display resolution | 0.1°C | ||
Temperature cycle setting | MAX 9999 cycles | ||
Ancillary Functions | Operation reservation function, zone selection function, operation start selection function, de-icing operation function, end state selection function, manual operation function, total time meter | ||
Temperature sensor | Pt100 (JIS C1604) | ||
Refrigeration method | Low-temperature tank Fully-enclosed, water-cooled, dual-refrigeration system Recovery unit Totally enclosed water-cooled single-stage refrigeration system | Low-temperature tank Scroll-type water-cooled dual-refrigeration system Recovery unit Totally enclosed water-cooled single-stage refrigeration system | |
refrigerant | Low temperature tank R-404A and R-23 Recovery unit R-404A | ||
heater | sheathed heater (SUS316L) | ||
ventilator | Propeller Type Agitator | ||
Damper drive unit | Air cylinders (2 vertical, 1 horizontal) | Air cylinders (4 vertical, 1 horizontal) | |
Brine Liquid Management System | Built-in vapor recovery device and automatic liquid feeder | ||
safeguard | Leakage circuit breaker, agitator current error, hot tank temperature error, cold tank temperature error, refrigerator 1 error, refrigerator 2 error, refrigerator for recovery unit error, air pressure error, liquid shortage, drive unit error, cooling water error, emergency stop switch | ||
power (button on TV, etc.) | AC200V 3-phase 50/60Hz Voltage fluctuation tolerance ±10% or less Noise level 3kV or less, noise period 1μsec or more, frequency fluctuation tolerance ±5% or less | ||
Maximum power consumption (kVA) | 10 | 16 | 18 |
Rated current (A) | 43 | 47 | 54 |
Primary-side electrical work | AC200V 3-phase 50/60Hz Voltage fluctuation tolerance ±10% or less Noise level 3kV or less, noise period 1μsec or more, frequency fluctuation tolerance ±5% or less | ||
Primary air construction | 0.4 to 0.7 MPa (dry air) External φ6mm hose connection | ||
coolant | Water volume 55 liter/min (water temperature 25℃) Bore diameter 20A | Water volume 70 liter/min(water temperature 25℃) Bore 25A | Water volume 100 liter/min (water temperature 25℃) Bore 32A |
drainage | Waste water condensed in refrigerator: 15A Separated waste water from recovered liquid: 15A Atmospheric release | ||
Standard accessories | Stainless steel (SUS304) sample basket (with temperature sensor T thermocouple), hose for drainage, hose for draining liquid, caster, instruction manual, warranty card, inspection certificate | ||