operability
Standard specifications include 5 operation patterns
In addition to constant value operation and auto start function operation, 2-step pattern operation 1-3 are standard specifications. Each of these pattern operations can be set to repeat, and one-touch operation is all that is required for the second and subsequent operations.
- Regular operation (auto-stop operation)Operation: Constant temperature and constant time.
- Auto start operationAfter waiting for a certain period of time, it operates at a certain temperature and for a certain period of time.
- Pattern 1 operation: Operates at two different temperatures and times (not including temperature change time).
- Pattern 2 operation: Operates at two different temperatures and times (including temperature change time).
- Pattern 3 operation (temperature gradient operation)Operation with two types of target temperatures and temperature change times.
Overshoot eliminated when temperature rises
Unique temperature control PID control system enables precise control at any temperature setting. Overshooting during temperature rise is prevented and control performance is improved.
Easy setup and operation
- interactive interaction
Setting items required for operation are displayed sequentially in an interactive format by pressing the menu key, and can be easily set by simply entering numerical values. - Intuitive Graphics
The control panel has a setting value input section on the right, and images connected by numerical values and symbols are collectively displayed; LEDs and pull-out lines make operation easy to understand and intuitive. - Japanese display
All displayed characters are in Japanese for ease of understanding. - simple operation
Four basic keys can be used to make settings with a few movements. Help display and guides make the interface easy to use even for beginners.
Setting value always displayed
The set temperature and remaining operation time are always displayed, allowing the current operation status to be checked at a glance.
Lightweight operation
Recessed operation keys, menu-driven setting items, and a bright, easy-to-read digital display provide a comfortable and enhanced operating feel.
Full range of included features
It is equipped with a full range of accessory functions, including a power failure recovery protection function, heater output monitoring function, and test termination output and trouble output.
- during a power outage: The set value and total time are backed up and not automatically restored (display alternates between the set temperature and remaining time).
- Heater output: Digital display of heater output during operation.
safety
Abnormal judgment detection step
HISPEC monitors multiple temperature changes in the tank, and in the event of abnormal temperatures, multi-layered safety controls are activated, shutting off the heater circuit as necessary to ensure safety.
- A/D conversion
- CPU Judgment
- Temperature Setting Comparison
- Over-rise and over-low judgment
- Heater circuit ON/OFF control
Response to changes in tank temperature
The HISPEC series is equipped with a threefold over-temperature protection mechanism to prevent abnormal temperature rises in the tank. First, a soft-controlled over-temperature limiter is activated, and if for some reason it is not activated, a hard-controlled over-temperature limiter is activated next. If the over-temperature protection still fails to stop, a thermal fuse finally interrupts the circuit to ensure safety in a multilayered manner.
Temperature condition | safety device | operation | display |
Thermal cutoff operating temperature exceeded | thermal fuse | Forced heater circuit break | nashi (Pyrus pyrifolia, esp. var. culta) |
Exceeds max. specification temperature +50°C | Over-temperature protector 1 (ETACOM: Hard) | Heater circuit OFF | Abnormality lamp + Er01 light + Alarm sound |
Upper setpoint temperature exceeded | Over-temperature protector 2 (ETACOM: Soft) | Heater circuit OFF | Abnormality lamp + Er02 light + Alarm sound |
Exceeds temperature setting by +10°C | Built-in temperature controller (etacom) | Continuous Drive | AL1 lighting + alarm sound (intermittent sound) |
Set temperature | Built-in temperature controller (etacom) | Continuous Drive | AL2 lighting + alarm sound (intermittent sound) |
Lower limit set temperature | Built-in temperature controller (etacom) | Heater circuit OFF | Abnormal lamp + heater circuit OFF + alarm sound |
series structure
How to read the model
S
type
Horizontal:S
Vertical:HS
Safety specification for solvent drying:HG
High temperature:EHT
/
3
Max. temperature
2: 200°C
3: 300°C
/
20
Internal dimensions: Horizontal (WxHxD mm)
10 : 450×450×450
20:600×600×600
40:800×1000×800
50: 1000 x 1000 x 800
Internal dimensions: Vertical (WxHxD mm)
40 : 600 x 900 x 600
50 : 450×450×450
60 : 450×450×450
/
sadist
type
sadistWith explosion vent
kForced exhaust type
NNon-oxidation specification
indecentHigh temperature specification
HISPEC horizontal type
The horizontal model is the standard model of the HISPEC series. It heats a large sample chamber efficiently and combines stable temperature distribution performance with ease of operation.
Highly accurate temperature distribution performance by eliminating overshoot
In addition to an in-vessel structure that takes advantage of many years of chamber technology, the use of a propeller fan with high air volume greatly improves circulation performance. Excellent temperature distribution performance of ±1.5°C (±2.0°C for #40 type and above) and high stability are realized.
Functions can be switched according to purpose
The hot air circulation function and the unidirectional hot air exhaust function can be used separately by switching the ventilation damper on the front, lower left or upper left side of the main unit. This function can be flexibly adapted according to the sample and purpose.
Simple maintenance zones
The number of parts used has been thoroughly reduced to improve reliability in long-term operation. As a result, the number of maintenance requirements has been greatly reduced, and the reliability of the entire system has been further enhanced.
Unitization of heater section realizes cost reduction
ETAC's uniquely designed heater section is commonized with other models by unitization. Cost reduction and standardization of assembly processes are realized.
specification sheet
model (e.g. of a vehicle) | HISPEC Horizontal 200°C series | |||
(technical) specification | HT210 | HT220 | HT240 | HT250 |
Circulation method | Forced hot air circulation and ventilation system | |||
Temperature range: | (room temperature +20°C to 200°C) | |||
Temperature distribution | ±1.5°C (at 200°C) | ±2.0°C (at 200°C) | ||
Time to reach max. temperature | Within 40 minutes (room temperature to 200°C) | Within 45 minutes (room temperature to 200°C) | ||
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | |||
Setting and display resolution | 1°C | |||
Setting and display accuracy | 0.5%(F.S)+1°C/digit | |||
sensor | K thermocouple | |||
Exterior Material | Cold-rolled steel sheet, baked finish | |||
Tank interior material | Stainless steel plate (SUS304) | |||
heat insulating material | glass wool | |||
Heating Heater | Nichrome Strip Heaters | |||
electric motor | Single-phase electric motor | |||
fan | propeller fan | |||
damper | Switchable between circulation and ventilation | |||
exhaust port | diameter (units assumed, esp. mm) | |||
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure return protection, momentary power failure retention mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, over-temperature alarm at set temperature +10°C, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse Door return protection mechanism | |||
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function | |||
power (button on TV, etc.) | AC200V single phase 50/60Hz | AC200V 3 phase 50/60Hz | ||
Maximum load current (A) | 10 | 13 | 14 | 17 |
External dimensions (W x H x D mm) | 1040 x 820 x 670 | 1190 x 970 x 820 | 1460 x 1570 x 1030 | 1660 x 1570 x 1030 |
Internal dimensions (W x H x D mm) | 450 x 450 x 450 | 600 x 600 x 600 | 800×1000×800 | 1000×1000×800 |
Volume (liter) | 91 | 216 | 640 | 800 |
Mass (kg) | 85 | 120 | 325 | 345 |
handling instructions | one copy | |||
volume | 2 shelves, 4 shelf supports | |||
Minimum mounting pitch | 50mm | |||
Total number of shelves | 8 | 11 | 19 | |
power cord | Three-core cabtyre cord (with grounding) 2m | – | ||
warranty | Separate mailing | |||
model (e.g. of a vehicle) | HISPEC Horizontal Type 300°C Series | |||
(technical) specification | HT310 | HT320 | HT340 | HT350 |
Circulation method | Forced hot air circulation and ventilation system | |||
Temperature range: | (Room temperature +20°C to 300°C) | |||
Temperature distribution | ±2.5°C (at 300°C) | ±3.0°C (at 300°C) | ||
Time to reach max. temperature | Within 50 minutes (room temperature to 300°C) | Within 60 minutes (room temperature to 300°C) | ||
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | |||
Setting and display resolution | 1°C | |||
Setting and display accuracy | 0.5%(F.S)+1°C/digit | |||
sensor | K thermocouple | |||
Exterior Material | Cold-rolled steel sheet, baked finish | |||
Tank interior material | Stainless steel plate (SUS304) | |||
heat insulating material | glass wool | |||
Heating Heater | Nichrome Strip Heaters | |||
electric motor | Single-phase electric motor | |||
fan | propeller fan | |||
damper | Switchable between circulation and ventilation | |||
exhaust port | diameter (units assumed, esp. mm) | |||
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure return protection, momentary power failure retention mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, over-temperature alarm at set temperature +10°C, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse Door return protection mechanism | |||
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function | |||
power (button on TV, etc.) | AC200V single phase 50/60Hz | AC200V 3 phase 50/60Hz | ||
Maximum load current (A) | 13 | 16 | 17 | 19 |
External dimensions (W x H x D mm) | 1040 x 820 x 670 | 1190 x 970 x 820 | 1460 x 1570 x 1030 | 1660 x 1570 x 1030 |
Internal dimensions (W x H x D mm) | 450 x 450 x 450 | 600 x 600 x 600 | 800×1000×800 | 1000×1000×800 |
Volume (liter) | 91 | 216 | 640 | 800 |
Mass (kg) | 85 | 120 | 325 | 345 |
handling instructions | one copy | |||
volume | 2 shelves, 4 shelf supports | |||
Minimum mounting pitch | 50mm | |||
Total number of shelves | 8 | 11 | 19 | |
power cord | Three-core cabtyre cord (with grounding) 2m | – | ||
warranty | Separate mailing | |||
HISPEC Vertical
The space-saving design of this vertical model enables processing of a large number of samples while reducing the footprint. Forced circulation by a sirocco fan provides sufficient uniform heating even when the distance between samples is relatively narrow.
Space-saving design
The main unit is vertically oriented and the exhaust port is located on the ceiling. This saves space in the installation area.
Capable of heat treating many samples at the same time
Hot air circulation is forced by a sirocco fan, which ensures sufficient circulation efficiency even when the distance between specimens is relatively small, allowing heat treatment of many specimens at the same time.
Heater unit with excellent durability
Sheathed heaters are standard equipment, enabling safe and long-term continuous use even when a large number of samples are loaded.
Dial damper knob
The damper knob is a dial type, allowing the degree of opening to be determined at a glance. It is also equipped with a lock mechanism to prevent accidental operation.
Adoption of heat-resistant decorative frames for control equipment
A heat-resistant decorative frame is used to prevent adverse effects on control equipment caused by opening and closing the door, even during high-temperature operation.
Improved safety when opening and closing doors during operation
Opening the door during high-temperature operation activates the door limit switch and stops the circulation fan, preventing workers from being blown by the hot air.
specification sheet
model (e.g. of a vehicle) | HISPEC Vertical 200°C series | ||
(technical) specification | HS240 | HS250 | HS260 |
Circulation method | Forced hot air circulation and ventilation system | ||
Temperature range | (room temperature +20°C) to 200°C | ||
Temperature distribution | ±2.0°C (at 200°C) | ||
Time to reach max. temperature | Within 40 minutes (room temperature to 200°C) | ||
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | ||
Setting and display resolution | 1°C | ||
Setting and display accuracy | 0.5%(F.S)+1°C/digit | ||
sensor | K thermocouple | ||
Exterior Material | Cold-rolled steel sheet, baked finish | ||
Tank interior material | Stainless steel plate (SUS304) | ||
heat insulating material | glass wool | ||
Heating Heater | sheath heater | ||
electric motor | Single-phase electric motor | ||
fan | sirocco fan | ||
damper | Switchable between circulation and ventilation | ||
exhaust port | Outer diameter of the mating part φ89 | ||
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in ETACOM), separate circuit over-temperature protection device, blower thermal, power failure recovery protection, momentary power failure retention mechanism, alarm, lower limit alarm, interlock terminal, error display, over-temperature alarm (+10°C), self-diagnostic function (sensor disconnection, heater disconnection, door-close alarm), temperature fuse, door-open temporary stop mechanism Thermal fuse, door return prevention mechanism, pause mechanism when door is open | ||
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function | ||
power (button on TV, etc.) | AC200V 3 phase 50/60Hz | ||
Maximum load current (A) | 15 | 17 | 18 |
External dimensions (W x H x D mm) | 770 x 1581 x 910 | 770 x 1881 x 910 | 970 x 1881 x 910 |
Internal dimensions (W x H x D mm) | 600×900×600 | 600×1200×600 | 800 x 1200 x 600 |
Volume (liter) | 324 | 432 | 576 |
Mass (kg) | 185 | 205 | 275 |
handling instructions | 1 copy | ||
volume | 2 shelves, 4 shelf supports | ||
Minimum mounting pitch | 50mm | ||
Total number of shelves | 15 | 21 | |
Casters & Adjusters | standard equipment | ||
warranty | Separate mailing | ||
HISPEC S type with explosion vent
This high-temperature unit with explosion vent is suitable for environments where flammable gas (vapor) is generated during drying. In the event of an explosion, the pressure is safely released from the exhaust vent in the ceiling to protect workers and the surrounding environment. However, it cannot be used when large amounts of flammable vapors are generated.
Equipped with vents to exhaust wind pressure completely
A pressure-relief opening (vent) is installed in the ceiling. Venting is safe in the event of an explosion.
Safety-conscious structure
The exhaust section is securely covered to a height of 1.8 m above the floor. The door perimeter is equipped with a locking mechanism that makes it easy to open and close while maintaining sufficient strength.
Superior operability
Easy operation from temperature setting to start of operation provides an intuitive operating feel common to the series.
specification sheet
model (e.g. of a vehicle) | HISPEC Horizontal 200°C series | |||
(technical) specification | HT210S | HT220S | HT240S | HT250S |
Circulation method | Forced hot air circulation and ventilation system | |||
Temperature range: | (room temperature +20°C to 200°C) | |||
Temperature distribution | ±1.5°C (at 200°C) | ±3.0°C (at 200°C) | ||
Time to reach max. temperature | Within 40 minutes (room temperature to 200°C) | Within 50 minutes (room temperature to 200°C) | ||
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | |||
Setting and display resolution | 1°C | |||
Setting and display accuracy | 0.5%(F.S)+1°C/digit | |||
sensor | K thermocouple | |||
Exterior Material | Cold-rolled steel sheet, baked finish | |||
Tank interior material | Stainless steel plate (SUS304) | |||
heat insulating material | glass wool | |||
Heating Heater | sheath heater | |||
electric motor | Single-phase electric motor | |||
fan | propeller fan | |||
damper | Switchable between circulation and ventilation | |||
exhaust port | diameter (units assumed, esp. mm) | |||
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure return protection, momentary power failure holding mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, set temperature +10°C over-temperature alarm, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse Explosion vent | |||
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function | |||
power (button on TV, etc.) | AC200V single phase 50/60Hz | AC200V 3 phase 50/60Hz | ||
Maximum load current (A) | 10 | 13 | 14 | 17 |
External dimensions (W x H x D mm) | 1040 x 1260 x 680 | 1190 x 1260 x 830 | 1460 x 1815 x 1030 | 1660 x 1895 x 1030 |
Internal dimensions (W x H x D mm) | 450 x 450 x 450 | 600 x 600 x 600 | 800×1000×800 | 1000×1000×800 |
Volume (liter) | 91 | 216 | 640 | 800 |
Mass (kg) | 90 | 125 | 290 | 310 |
handling instructions | one copy | |||
volume | 2 shelves, 4 shelf supports | |||
Minimum mounting pitch | 50mm | |||
Total number of shelves | 8 | 11 | 19 | |
Vents | 1 formula | |||
power cord | Three-core cabtyre cord (with grounding) 2m | – | ||
warranty | Separate mailing | |||
model (e.g. of a vehicle) | HISPEC Horizontal Type 300°C Series | |||
(technical) specification | HT310S | HT320S | HT340S | HT350S |
Circulation method | Forced hot air circulation and ventilation system | |||
Temperature range: | (Room temperature +20°C to 300°C) | |||
Temperature distribution | ±2.5°C (at 300°C) | ±4.0°C (at 300°C) | ||
Time to reach max. temperature | Within 50 minutes (room temperature to 300°C) | Within 60 minutes (room temperature to 300°C) | ||
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | |||
Setting and display resolution | 1°C | |||
Setting and display accuracy | 0.5%(F.S)+1°C/digit | |||
sensor | K thermocouple | |||
Exterior Material | Cold-rolled steel sheet, baked finish | |||
Tank interior material | Stainless steel plate (SUS304) | |||
heat insulating material | glass wool | |||
Heating Heater | sheath heater | |||
electric motor | Single-phase electric motor | |||
fan | propeller fan | |||
damper | Switchable between circulation and ventilation | |||
exhaust port | diameter (units assumed, esp. mm) | |||
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure return protection, momentary power failure holding mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, set temperature +10°C over-temperature alarm, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse Explosion vent | |||
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function | |||
power (button on TV, etc.) | AC200V single phase 50/60Hz | AC200V 3 phase 50/60Hz | ||
Maximum load current (A) | 15 | 17 | 19 | |
External dimensions (W x H x D mm) | 1040 x 1260 x 680 | 1190 x 1260 x 830 | 1460 x 1815 x 1030 | 1660 x 1895 x 1030 |
Internal dimensions (W x H x D mm) | 450 x 450 x 450 | 600 x 600 x 600 | 800×1000×800 | 1000×1000×800 |
Volume (liter) | 91 | 216 | 640 | 800 |
Mass (kg) | 90 | 125 | 290 | 310 |
handling instructions | one copy | |||
volume | 2 shelves, 4 shelf supports | |||
Minimum mounting pitch | 50mm | |||
Total number of shelves | 8 | 11 | 19 | |
Vents | 1 formula | |||
power cord | Three-core cabtyre cord (with grounding) 2m | – | ||
warranty | Separate mailing | |||
HISPEC K type Hot air forced exhaust type
This is a hot-air exhaust type high-temperature unit developed for heat treatment of secondary vulcanization as typified by silicone rubber. It uses a one-way complete exhaust system that efficiently discharges gases during vulcanization to the outside of the tank, thereby enhancing the safety of the work environment.
Discharge of vulcanized folded material to the outside
Fresh air induction and forced exhaust increases circulation efficiency, and treatment is performed without affecting vulcanized materials or external components.
Double swinging door for easy operation
Double-sliding structure reduces the opening radius to half. Improves productivity during high-temperature operations.
Easy-to-operate control functions
Auto start/stop/weight functions are standard, allowing easy operation with one-touch operation the next time.
specification sheet
model (e.g. of a vehicle) | HISPEC Horizontal type | |
(technical) specification | HT340K | HT350K |
Temperature range | (room temperature +20°C to 250°C) | |
Temperature distribution | ±5°C (at 250°C) | |
Time to reach max. temperature | Within 60 minutes (room temperature to 250°C) | |
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method | |
Setting and display resolution | 1°C | |
Setting and display accuracy | 0.5%(F.S)+1°C/digit | |
sensor | K thermocouple | |
Exterior Material | Cold-rolled steel sheet, baked finish | |
Tank interior material | Stainless steel plate (SUS304) | |
heat insulating material | glass wool | |
door | double-opening (e.g. door, window) | |
Heating Heater | Nichrome Strip Heaters | |
electric motor | Single-phase electric motor | |
fan | propeller fan | |
Air intake blower | Sirocco fan/motor type | |
Compact Blower | Motor shaft cooling blower | |
exhaust port | Ceiling section φ120 | |
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure return protection, momentary power failure retention mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, over-temperature alarm at set temperature +10°C, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse Door return protection mechanism | |
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step hold operation function, weight function | |
power (button on TV, etc.) | AC200V 3 phase 50/60Hz | |
Maximum load current (A) | 28 | |
External dimensions (W x H x D mm) | 1460 x 1780 x 1030 | 1660 x 1780 x 1030 |
Internal dimensions (W x H x D mm) | 800×1000×800 | 1000×1000×800 |
Volume (liter) | 640 | 800 |
Mass (kg) | 365 | 375 |
handling instructions | one copy | |
volume | 2 shelves, 4 shelf supports | |
Minimum mounting pitch | 50mm | |
Number of shelves | 19 | |
warranty | Separate mailing | |
HISPEC HG type High temperature tank with safety specifications for solvent drying
The HG type is a high-temperature unit with safety specifications for solvent drying, designed based on the concept that "safety takes precedence over everything else. It can safely perform heating and drying operations that involve the generation of flammable vapors, such as ceramic molding processes and drying after cleaning electronic components.
Adoption of forced exhaust system
A forced exhaust airflow system is used to blow air inside the tank. Flammable gases generated in the tank are constantly exhausted to the outside, preventing their concentration from rising.
Gas concentration alarm as standard equipment
Equipped with a high-temperature compatible gas concentration alarm to monitor the inside of the test chamber at all times. It is equipped with a safety function that enables operation only when the gas concentration is below the lower limit.
People-friendly safety design
The door structure is designed for ease of operation and safety, with a simple locking mechanism and safety interlock. Equipped with an uninterruptible power supply, the ventilation continues for a certain period of time even during a power failure.
Equipped with safety structures throughout
Various safety devices such as explosion vent mechanism are standard. Special inner tank structure design ensures high safety.
specification sheet
model (e.g. of a vehicle) | HISPEC HG220 |
Temperature range | (room temperature +20°C to 200°C) |
Temperature distribution | ±3.0°C (at 200°C) |
Time to reach max. temperature | Within 50 minutes (room temperature to 200°C) |
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method |
Setting and display resolution | 1°C |
Setting and display accuracy | 0.5%(F.S)+1°C/digit |
sensor | K thermocouple |
Exterior Material | Cold-rolled steel sheet, baked finish |
Tank interior material | SUS304 (All welding is done below the center of the inner tank) |
heat insulating material | glass wool |
Heating Heater | sheath heater |
electric motor | Single-phase electric motor |
fan | propeller fan |
exhaust port | 80mm dia. |
(engine) displacement | Approx. 1 m3/min or more |
electric motor | Single-phase electric motor |
fan | sirocco fan |
Explosion Vent | Vent for explosion pressure relief in tank |
Thermal overload protection device | Temperature control type For preventing over-temperature in the tank |
Heater over-temperature protector | Temperature control type For preventing heater ambient temperature from rising too high |
gas sensor | Operation stops at 30% of the lower explosion limit concentration in the tank |
UPS | Backup air supply fan and gas alarm for at least 5 minutes in case of power failure |
Other | Earth leakage circuit breaker, over-temperature protection device (built-in ETACOM), separate circuit over-temperature protection device, blower thermal, blower thermal for air supply, thermal fuse, heating delay mechanism, fan stop delay mechanism, external alarm terminal, abnormality display, lower limit set temperature alarm, set temperature +10°C over-temperature alarm, door limit switch, self-diagnosis function (sensor disconnection, heater disconnection), alarm buzzer |
Ancillary Functions | Memory backup (approx. 10 years), last step hold operation function, auto start, auto stop, programmed operation function (2 steps x 3 patterns) |
power (button on TV, etc.) | AC200V 3 phase 50/60Hz |
Maximum load current (A) | 19 |
External dimensions (W x H x D mm) | 1215 x 1915 x 980 |
Internal dimensions (W x H x D mm) | 600 x 600 x 600 |
Volume (liter) | 216 |
Mass (kg) | 200 |
handling instructions | one copy |
volume | 2 shelves, 4 shelf supports |
Minimum mounting pitch | 50mm |
Number of shelves | 11 |
warranty | Separate mailing |
HISPEC N type Non-oxidation incubator
Heat treatment of components with copper or silver electrodes requires careful attention to the relationship between temperature and oxide film. HISPEC N type guarantees a residual oxygen concentration of 100ppm or less, realizing a heat treatment environment with minimal oxidation. It enables high-performance and low-cost non-oxidation testing, contributing to worker safety and productivity improvement.
Guaranteed residual oxygen concentration of 100ppm or less
It can stably reproduce residual oxygen concentrations of 100 ppm or less while reducing the amount of N₂ gas used to 1/5 of the conventional level.
Improved temperature lowering mechanism
Ability to lower temperature from 250°C with the door closed is 1.5 times faster than conventional units. Expands the workable area. If further shortening is required, a water cooling mechanism (optional) is available.
Temperature uniformity
Temperature distribution performance of ±3.5°C (at 250°C setting) is guaranteed. Improves heat treatment yield.
specification sheet
Product name | Non-oxidation incubator |
model (e.g. of a vehicle) | HT320N |
Circulation method | Forced hot air circulation system |
Temperature range | (room temperature +40°C) to 250°C |
Temperature distribution | ±3.5°C (at 250°C) |
Temperature rise time | Within 60 minutes (room temperature to 250°C) |
residual oxygen concentration | Less than 100ppm |
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method |
Setting and display resolution | 1°C |
Setting and display accuracy | 0.5%(F.S)+1.0°C/digit |
sensor | K thermocouple |
Exterior Material | Cold-rolled steel sheet, baked finish |
Tank interior material | Stainless steel plate, all welded construction |
heat insulating material | glass wool |
Heating Heater | sheath heater |
ventilator | Single-phase motor AC200V single-phase 200W |
fan | Stainless steel propeller fan φ300 |
N₂ inlet | 3/8B Socket |
N₂ exhaust port | 3/4B hose nipple |
N₂ pressure gauge | 0.6MPa (6kgf/cm2) with pressure adjustment valve |
N₂ flowmeter | 0-50 liter/min |
cooling mechanism | Air forced cooling (at end of operation) |
safeguard | Earth leakage circuit breaker, over-temperature protector (built-in etacom), separate circuit over-temperature protector, blower thermal, power failure return protection, momentary power failure holding mechanism, alarm, lower limit alarm, abnormal display, over-temperature alarm at set temperature +10°C, door return prevention mechanism, external alarm terminal, self-diagnostic function (sensor disconnection, heater disconnection), control circuit and heater circuit circuit circuit breaker |
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (20 steps x 3 patterns), last step hold operation function |
option | N₂ flowmeter with alarm, primary side regulator, oxygen analyzer (% level), oxygen analyzer (ppm level), abnormal pressure detection, cooling fan linked to time signal, processing end buzzer, alarm buzzer, rotating lamp, rotating lamp with buzzer, calendar timer, total time meter, shelf board, load-bearing shelf board, automatic power failure recovery function, test end Terminal, body specified color, RS485 interface, program function, remote function, time signal, weight function, temperature recorder, power cable (3.5mm 24-core/3m, 5m, 10m) |
accessory | Instruction manual, 2 shelves, 4 shelf supports |
External dimensions (W x H x D mm) | 1330 x 1645 x 900 |
Internal dimensions (W x H x D mm) | 600 x 600 x 600 |
Casters & Adjusters | standard equipment |
power (button on TV, etc.) | AC200V 3 phase 50/60Hz Power fluctuation tolerance within ±10 |
Maximum load current (A) | 13 |
Volume (liter) | 216 |
Mass (kg) | 160 |
HISPEC H type High-temperature unit
This model is suitable for heat resistance tests and high temperature life tests of inorganic materials such as ceramics and various metals, and engineering plastics at MAX500℃. It can be operated safely and precisely even in high temperature ranges.
High performance and high precision
Reaches MAX500°C in about 60 to 70 minutes. Unique PID control suppresses overshoot and provides excellent temperature distribution inside the tank.
Designed with safety in mind
Triple safety mechanism against abnormal temperatures (over-temperature limiter, upper limit setting, and thermal fuse). Door lock mechanism and highly sealed structure with double packing for long time use.
Easy-to-use controller
The HT/HS series shares the same operation system with the HT/HS series, with 2-step programmed operation, auto-start/auto-stop, and each safety device operation display as standard features. Intuitive operation.
specification sheet
model (e.g. of a vehicle) | EHT-2H |
Circulation method | Forced hot air circulation and ventilation system |
Temperature range: | (room temperature +40°C) to 500°C |
Temperature distribution | ±5.0°C (at 500°C) |
Time to reach max. temperature | Within 90 minutes (room temperature to 500°C) |
Control method | Microcomputer-controlled temperature controller (ETACOM), PID control method |
Setting and display resolution | 1°C |
Setting and display accuracy | 0.5%(F.S)+1.0°C/digit |
sensor | K thermocouple |
Exterior Material | Cold-rolled steel sheet, baked finish |
Tank interior material | Stainless steel plate (SUS304) |
heat insulating material | Heat-resistant glass wool |
Heating Heater | Nichrome Strip Heaters |
electric motor | AC200V 3 phase 400W |
fan | propeller fan |
damper | Switchable between circulation and ventilation |
exhaust port | φ82 (with exhaust duct) |
safety device | Earth leakage circuit breaker, over-temperature protection device (built-in etacom), separate circuit over-temperature protection device, blower thermal, power failure recovery protection, momentary power failure retention mechanism, alarm, lower limit set temperature alarm, interlock terminal, error indication, set temperature +10°C over-temperature alarm, self-diagnostic function (sensor disconnection, heater disconnection), thermal fuse |
Ancillary Functions | Memory backup (approx. 10 years), auto start, auto stop, programmed operation function (2 steps x 3 patterns), last step holding operation function |
power (button on TV, etc.) | AC200V 3 phase 50/60Hz |
Maximum load current (A) | 18 |
External dimensions (W x H x D mm) | 1440 x 1175 x 1115 (including exhaust duct) |
Internal dimensions (W x H x D mm) | 600 x 600 x 600 |
Volume (liter) | 216 |
Mass (kg) | 280 |
handling instructions | one copy |
volume | 2 shelves, 4 shelf supports |
Minimum mounting pitch | 50mm |
Number of shelves | 11 |
warranty | one copy |
option
A wide range of options to enable optimal specifications according to the purpose of use.
Over-temperature protector (temperature control type)
Used when quadruple safety or heater room temperature control is required in addition to the standard safety mechanism.
sheath heater
SUS304 for horizontal type, reinforced corrosion resistance and red-hot prevention, standard for HS/S/HG/N types.
Variable wind speed mechanism
Fan speed can be adjusted for testing under low air velocity conditions such as film drying.
Automatic damper
Automatic control of damper opening in conjunction with test end output, trouble output, and time signal.
Exhaust Duct
Effective for outdoor exhaust when rear exhaust is to be avoided. Optimizes the exhaust route to suit the environment and the line of flow.
Simple N₂ displacement mechanism
For treatment of specimens that do not tolerate oxidation. Simplified non-oxidizing atmosphere.
External alarm terminal / Test end terminal
Outputs abnormalities and termination to external equipment (standard for N and HG types). Contact rating of the test termination terminal is AC250V 3A.
buzzer switch
Audible notification of the end of the test or the occurrence of a problem.
rotation warning light
Linked to trouble output and visualizes abnormalities with a rotating flash (models with buzzer available).
emergency stop switch
In the event of an emergency, the breaker interrupts the current and immediately stops operation.
calendar timer
AM/PM for weekly automatic operation schedule.
paperless recorder
Can be saved to CF card and imported to PC with included software. Compatible with percussion recorder.
Observation window (with in-tank light)
Made of heat-resistant tempered glass; meets 200°C specifications and improves visibility during operation.
integration timer
The time used can be totaled and utilized for maintenance planning and operation analysis.
Trestle (assembled type)/Angle type trestle
Adjustable height for easy work. Angle type is equipped with casters for easy carrying in and installation.
Shelf / Load-bearing shelf board
Standard 15 kg and load capacity 50 kg (equally distributed). Shelf pitch is 50mm for flexible adjustment.
Racks & Rack Carts
Production transport fixture. Racks and carts are wheeled for ease of operation.
Measuring port (φ25/φ50/φ100)
Measurement port used to take out lead wires and sensors.
RS485 Interface
Supports remote monitoring and data collection via external communication (9600 bps/half-duplex/8N1).
Double swinging door (for horizontal type)
Reduced door turning radius saves space and improves workability for loading and unloading.
Recorder output terminal
K thermocouple signals can be output to an external recorder.
Casters & Adjusters
Integrated legs for both movement and leveling, standard on HS/N types.
Automatic Power-Outage Recovery Function
Automatically resumes operation after power failure. Supports unlimited time restoration.
Remote Functions
Operation start/stop can be remotely controlled by an external signal (remote terminal on rear panel).
Voltage application terminal
Voltage can be applied to the test object under test in high temperature operation tests, load life tests, etc.
time signal
Two contact points are activated at any step of pattern operation (AC250V 3A).
Program specifications (max. 20 steps)
Supports programmed operation of up to 20 steps x 3 patterns (N type standard).
weight function
Excluding the temperature rise time, only the maintenance time is counted (valid for constant value operation/standard for K type).
2-step metal fittings
Special fittings for loading horizontal type in two levels, top and bottom.
Floor load-bearing reinforcement
The tank floor is reinforced to support heat treatment of heavy materials.
Gear aging mechanism
This is a dedicated mechanism that is individually designed for each application.
Body paint color Specified
The body color can be specified to match the surrounding equipment and brand standards.