Basic performance
- Inside dimensionsW600 x H750 x D600 mm - W1000 x H1000 x D800 mm
- content by volumeL : 270 to 800 L
- Temperature range:-30°C to +100°C (+150°C specification available)
- Humidity range:20%RH~98%RH
Product Features
Long-term humidification test support
By connecting an additional water supply tank (optional), continuous operation for 10 days to a month is possible, reducing the risk of interruption during equipment operation and reducing the need to refill humidified water.
Dew condensation prevention function
When the temperature and humidity rise, the "humidification delay operation" suppresses the rate of change and starts humidification after reaching the set temperature, and the "normal temperature standby operation" returns the inside of the chamber to 23°C and stands by after the test is completed, preventing condensation. These functions provide a safe design that protects specimens from condensation and other problems.
High durability cooling coil
The cooling coil made of a new material greatly reduces corrosion caused by organic acid substances and improves the reliability of the refrigeration circuit. Furthermore, by selecting SUS cooling coils (optional), which have superior corrosion resistance, stable performance can be achieved even in more severe environments.
operability
Main unit controller
A simple key-operated controller is standard.
preset function
Frequently used test conditions can be easily preset, up to 26 steps x 9 patterns can be set, and up to 99 repeats are supported.
Smart Connect (optional)
Smart Connect, which allows operation, monitoring, and data downloading from a PC or tablet, is available.
Operation screen
Provides an easy-to-understand operation screen that displays trouble conditions and graphs program input.
Large observation window (optional)
Large, easy-to-see windows can be added (e.g., FX320).
Adjustable shelves
Installation interval of 43 mm allows flexible adjustment to suit the EUT.
Dust filter
Interchangeable from the front, allowing for flexibility in the installation environment.
Highly maintainable inner tank
Made of SUS304 for durability and easy daily cleaning.
Humidified water sterilization lamp (optional)
UV irradiation inhibits bacterial growth and prevents problems before they occur.
Standard measurement hole
A 50 mm dia. measuring hole is provided on the left side as standard equipment.
option
non-frosting device
Expands the range in which frost does not adhere to the cooling coils, enabling continuous testing for extended periods of time without defrosting operation. Ensures stability during low-temperature operation.
150°C specification
In addition to the standard specifications, the upper limit of test temperature can be extended to 150°C. It is suitable for high-temperature tests and special applications.
Floor reinforcement (200 kg)
Option to reinforce the standard load capacity of 80 kg to 200 kg. Effective for handling heavy specimens.
Smart Controller (web substrate)
Direct access from a PC or tablet to set operating conditions, review history, and download data in CSV format. Improves manageability in a networked environment.
SUS cooling coil specifications
The cooling coil made of corrosion-resistant SUS provides durability even in tests that handle samples that generate organic acids and other substances.
Shelf board / Shelf support
3-mesh wire mesh specification adapts to testing of fine parts. Flexible layout is possible with a load capacity of 15 kg.
Load-bearing shelf board
Available in two types, one for 30 kg and the other for 50 kg, the system allows for safe installation of heavy objects.
Weipalace Observation Window
Triple heat-resistant glass with conductive finish enables fog-free observation even under low temperature and high humidity conditions.
over-temperature protector
In addition to the standard double-layer structure, a triple-layer structure provides safer protection for samples. This prevents abnormal temperature rises.
emergency stop switch
A safety device that can shut down the breaker and immediately stop chamber operation in the event of an emergency.
humidified water disinfection lamp
UV irradiation inhibits bacterial growth in humidified water. It also reduces wetness and water stains in the piping and improves maintainability.
Additional water tank
An additional 20L capacity tank can be connected to support long-term continuous operation. Reduces water supply labor.
Additional Cable Hole
Round φ50, square 100 x 30 mm, round φ120, etc. can be added. Supports external devices and wiring routing.
Plugs for various cable holes
Equipped with a soft-type silicone stopper and band packing for both sealing and operability.
Inner door with control hole
An observation window made of tempered glass is provided with an operation hole to allow internal operation while checking the test condition.
Water purifier (built-in type/separate type)
High purity pure water can be supplied by using ion exchange resin. Separately installed type for tap water can also be selected.
Paperless recorder (SD card storage)
Test data can be saved to SD card and imported to PC with dedicated software. Efficient recording without using paper.
Temperature/humidity output terminal (4-20 mA DC)
The temperature and humidity inside the tank can be output externally via analog signals and linked to other monitoring equipment.
RS485 interface
Supports remote operation and data communication, facilitating system integration.
Signal tower, rotating warning lights
The system notifies the user of the condition under test by color and lighting, and emits a highly visible warning in the event of an abnormality.
Butt for chamber body (leakage prevention)
The floor batts catch water leaks in case of emergency, and specifications with leakage sensors are also available.
Automatic Power-Outage Recovery Function
Automatically resumes operation upon return from a power outage, minimizing interruptions to the test.
Main body fixing metal fittings
Fittings for anchoring to the floor of the installation site with anchor bolts, effective against earthquakes and vibrations.
Power cable (3m or 5m)
4-wire cabtyre cable can be selected. Can be used in accordance with the installation environment on site.
Inspection report, calibration certificate, traceability chart
Provides a variety of documents to prove product performance. This is useful for quality assurance and audit support.
Humidified water direct supply specification (for pure water)
Humidified water can be supplied directly when pure water piping facilities are available.
Humidified water direct supply specification with purifier (for tap water)
In the absence of pure water facilities, this system supplies tap water by converting it to pure water.
Humidified water monitoring function
An alarm notifies the user if the quality of the humidified water exceeds standards, ensuring the reliability of the test.
specification sheet
model (e.g. of a vehicle) | FX310 | FX320 | FX330 | FL310 | FL320 | FL330 |
Temperature control range | -30°C to +100°C | |||||
Humidity control range | 20% RH to 98% RH | – | ||||
Temperature rise time | Within 45 minutes From -30°C to +100°C | Within 50 minutes From -30°C to +100°C | Within 45 minutes From -30°C to +100°C | Within 45 minutes From -30°C to +100°C | Within 50 minutes From -30°C to +100°C | Within 45 minutes From -30°C to +100°C |
Temperature fall time | Within 40 minutes +20°C to -20°C | Within 35 minutes +20°C to -20°C | Within 45 minutes +20°C to -20°C | Within 40 minutes +20°C to -20°C | Within 35 minutes +20°C to -20°C | Within 45 minutes +20°C to -20°C |
Temperature change rate (rise) | 3.5°C/min From -17.0°C to +87.0°C | 3.0°C/min From -17.0°C to +87.0°C | 4.0°C/min From -17.0°C to +87.0°C | 3.5°C/min From -17.0°C to +87.0°C | 3.0°C/min From -17.0°C to +87.0°C | 4.0°C/min From -17.0°C to +87.0°C |
Temperature change rate (descending) | 2.5°C/min From +87.0°C to -17.0°C | 3.0°C/min From +87.0°C to -17.0°C | 1.0°C/min From +87.0°C to -17.0°C | 2.5°C/min From +87.0°C to -17.0°C | 3.0°C/min From +87.0°C to -17.0°C | 1.0°C/min From +87.0°C to -17.0°C |
Temperature and humidity fluctuations | ±0.3°C/±2.5%RH | – | ||||
Temperature fluctuation | – | ±0.3°C | ||||
Temperature and humidity gradient | 3.0°C during temperature operation|1.5°C/5.0%RH during temperature/humidity operation | – | ||||
temperature gradient | – | 3.0°C | ||||
Space temperature and humidity deviation | 2.5°C during temperature operation|1.5°C/5.0%RH during temperature/humidity operation | – | ||||
Space temperature deviation | – | 2.5°C | ||||
Operable ambient temperature range | +5°C to +40°C (The minimum temperature may not be maintained if the temperature at the installation site exceeds +35°C.) | |||||
Internal dimensions (W x H x D mm) | 600×750×600 | 700×950×700 | 1000×1000×800 | 600×750×600 | 700×950×700 | 1000×1000×800 |
Content (liter) | 270 | 465 | 800 | 270 | 465 | 800 |
External dimensions (W x H x D mm) | 880 x 1653 x 1195 | 980 x 1853 x 1295 | 1280 x 1914 x 1395 | 880 x 1653 x 1195 | 980 x 1853 x 1295 | 1280 x 1914 x 1395 |
Mass (kg) | 305 | 380 | 495 | 295 | 370 | 485 |
Exterior Steel Plates | Steel plate with baked finish | |||||
Inner tank steel plate | Stainless steel plate (SUS304) | |||||
ventilator | sirocco fan | |||||
refrigerator | Air-cooled twin rotary chillers | |||||
refrigerant | R-448A | |||||
Heating Heater | Nichrome Strip Heaters | |||||
humidifier | sheathed heater (SUS316L) | – | ||||
cooler | Plate Fin Coil | |||||
Control Sensors | JIS C1604 3-wire resistance thermometer Pt100 | |||||
Control method | PID control | |||||
Display method | LED indication | |||||
Display resolution | 0.1°C/1%RH | 0.1°C | ||||
Setting method (standard) | Key-in method using main unit setting device Constant value operation, programmed operation – 9 programs can be saved (26 steps per program), upper and lower limit alarm function (temperature and humidity), operation status check monitor function | Key-in method using main unit setting device Constant value operation, programmed operation – 9 programs can be saved (26 steps per program), upper and lower limit alarm function (temperature), operation status check monitor function | ||||
Setting method (optional) | Smart Connect (optional) touch panel method, Smart Controller (optional) web browser method program operation – 99 programs can be stored (50 steps per program), program repeat function (up to 999 times, whole or partial setting possible) possible) | |||||
power (button on TV, etc.) | AC200V 3φ 50/60Hz Voltage fluctuation range ±10% Class D grounding | |||||
humidified water | Ion-exchanged water or distilled water (0.1 to 10 [μ S/cm]) | – | ||||
Maximum load current (A) | 24 | 29 | 32 | 19 | 24 | 30 |
Operating current (A) | 20 | 22 | 13 | 19 | ||
Exhaust heat (kW) | 3.3 to 3.8 | 3.4 to 4.9 | 3.3 to 3.8 | 3.3 to 3.8 | 3.4 to 4.9 | 3.3 to 3.8 |
safeguard | Main power leakage circuit breaker, fan motor overload relay, dryer overheat protector, refrigerator pressure switch, refrigerator overload relay, refrigerator overheat protector, primary power supply reverse phase detector, heating heater overload protection fuse, humidifier overload protection fuse, control circuit overload protection fuse, sensor disconnection detection function, temperature and humidity upper and lower limits Alarm function, power failure notification alarm, filter cleaning alarm, wick drying alarm, low water level alarm | Main power leakage circuit breaker, fan motor overload relay, refrigerator pressure switch, refrigerator overload relay, refrigerator overheat protector, primary power supply reverse phase detector, heating heater overload protection fuse, control circuit overload protection fuse, sensor disconnection detection function, temperature upper and lower limit alarm function, power failure notification alarm, filter cleaning Alarms | ||||
Ancillary Functions | Memory backup function, humidification delay operation function, normal temperature standby operation function, test end operation function, wait function, timer function, hold operation function, pause function, equipment operation status display, power failure return protection function, instantaneous power failure hold function, self-diagnostic function, auto start/timer function | Memory backup function, normal temperature standby operation function, test end operation function, wait function, timer function, hold operation function, pause function, equipment operation status display, power failure return protection function, instantaneous power failure hold function, self-diagnostic function, auto start/timer function | ||||
body (of a machine) | Circular measuring hole (φ50, 1 location on left side), caster adjuster, dust filter for refrigerator condenser | |||||
accessory | Instruction manual (1 set), ø50 silicon plug for measurement hole, water supply tank (FX310, FX320: 10 liters x 1 piece, FX330: 10 liters x 2 pieces), wick gauze (12 pieces), inspection certificate | Instruction manual (1 set), silicon plug for φ50 measurement hole, inspection certificate | ||||