WeissTechnikSpecimen Transfer type Cold Thermal Shock Chamber

ShockEvent

ShockEvent is an international standard-compliant thermal shock tester developed by WeissTechnik. It reproduces instantaneous temperature changes under a wide range of conditions from -80°C to +220°C by using a sample transfer system with a high-temperature chamber and a low-temperature chamber. Thermal stresses applied to electronic components and materials can be accurately evaluated. High load capacity, defrost-free operation, and energy-saving design ensure both efficiency and reliability. Furthermore, it complies with environmental regulations by using low GWP refrigerant R449A. With a wide variety of models and intuitive operability, this next-generation solution meets the testing needs of diverse industries such as automotive, electronics, and aerospace.

WeissTechnikindicates word or phrase being defined

Weiss Technik

Basic performance

  • Temperature range:: -80°C to +220°C
  • Test area travel timeWithin 10 seconds
  • Maximum load capacityMaximum weight: 100 kg (T/300 model)
  • Return Time to TemperatureWithin 12-15 minutes (under standard conditions)
  • High airtightness and corrosion resistanceStainless steel test chambers are used
Product Catalog
specification sheet

Product Features

Sample transfer type structure

By switching between high and low temperature chambers in a short period of time, specimens are subjected to instantaneous temperature changes to realize highly reproducible cold thermal shock testing.

Defrost-free operation

The unique mechanism enables continuous operation of more than 1,000 cycles, eliminating the need for test interruption due to defrosting.

Independent use of high and low temperature tanks

Each chamber can be used as an independent chamber and can be utilized not only for thermal shock testing but also for independent high or low temperature testing.

Robust construction

The test chamber is made of highly corrosion-resistant stainless steel to maintain stable performance even with rapid temperature changes and long-term operation. Product Features

operability

  • Large observation windowWidth: 355 mm x Height: 355 mm. all models are equipped as standard for high visibility.
  • Sample basket with meshSafety is improved by preventing specimens from falling.
  • access portReduces the load on the cable and maintains measurement accuracy.
  • 7-inch color TFT touch panelIntuitive operation screen with full Japanese support.
  • WEBSeason®.Remote access from smartphones and tablets, with chart preview function.
  • S!MPATI®.: Up to 99 test instruments can be centrally managed, and data can be documented and edited (optional).
  • CSV output: Test results can be saved on a USB memory stick and analyzed with common spreadsheet software.

Environmentally Friendly

  • Adoption of low GWP refrigerant R449AGlobal warming potential is reduced to about 1/3 that of conventional refrigerants.
  • Compliance with European F-Gas Regulations: Can be used safely even after 2030
  • economical modeEnergy savings of up to 37% is achieved by controlling the unused tanks without heating or cooling them.
  • CO2 emission reductionContributing to sustainability by reducing environmental impact

(product) line-up

ShockEvent offers five basic models, which can be selected according to sample size and test purpose.

specification sheet

View in PDF

model (e.g. of a vehicle)

T/60/V2

T/120/V2

T/120/V2/P

T/300/V2

T/300/V2/P

system

Two-zone method with vertical sample movement

High temperature exposure temperature range

+50°C to +220°C

Low temperature exposure range

-80°C to +70°C

Temperature fluctuation

±0.3K to ±1.0K

±1.0K

Space temperature deviation

±0.5K to ±2.0K

±2.0K

Preheat upper limit temperature

+220°C

Temperature rise time

Approx. 10 min.

From +50°C to +200°C

Approx. 12 min.

From +50°C to +200°C

Approx. 7 min.

From +50°C to +200°C

Approx. 14 min.

From +50°C to +200°C

Approx. 8 min.

From +50°C to +200°C

Temperature change rate (rise)

17K/min

14K/min

18K/min

11K/min

23K/min

Lower limit of pre-cooling temperature

-80°C

Temperature fall time

Approx. 70 min.

+70°C to -80°C

Approx. 41 min.

+70°C to -80°C

Approx. 33 min.

+70°C to -80°C

Approx. 45 min.

+70°C to -80°C

Approx. 26 min.

+70°C to -80°C

Temperature change rate (descending)

3.7K/min

6.3K/min

7.5K/min

5.0K/min

12K/min

Temperature change rate (rise)

3.2K/min

2.0K/min

2.0K/min

1.5K/min

1.5K/min

return condition

MIL-STD-883K 1010.9 Level D: IC4.5kg

MIL-STD-883K 1010.9 Level D: IC12kg

MIL-STD-883K 1010.9 Level D: IC20kg

MIL-STD-883K 1010.9 Level F: IC25kg

MIL-STD-883J 1010.9 Level F: IC50kg

return time

<15 min.

<15 min.

<12 min.

<15 min.

Test area travel time

<10 sec.

Allowable heat generation load

Max. approx. 2.0 kW

(-50°C to 0°C)

Max. approx. 6.0 kW

(-50°C to 0°C)

Max. approx. 8.0 kW

(-50°C to 0°C)

Operable ambient temperature range/relative humidity

+10 to +35°C/75%RH or less

External dimensions (W x H x D mm)

900 x 2330 x 2150

1000 x 2450 x 2530

1000×2450×2805

1320 x 2900 x 2370

1320 x 2900 x 2370

External dimensions of separately installed chiller unit (W x H x D mm)

800 x 1950 x 2100

1000 x 2080 x 2400

Minimum external dimensions when carried in (W x H x D mm)

900 x 1895 x 2220

1000 x 1985 x 2600

1000 x 1985 x 2875

1350 x 2250 x 2340

1350 x 2250 x 2340

Minimum external dimensions (W x H x D mm) when separately installed chiller unit is carried in

800 x 1920 x 1920

1000 x 2050 x 2330

Mass (kg)

Approximately 800

Approx. 1000

Approx. 1200

Approx. 1630 (950+680)

Approx. 1950(1100+850)

Test area dimensions (W x H x D mm)

380 x 370 x 430

470 x 410 x 650

770 x 610 x 650

Test area volume (liter)

60

120

300

Max. number of shelves

four flat objects (e.g. tickets, pieces of cloth, etc.)

five flat objects (e.g. tickets, pieces of cloth, etc.)

eight flat objects (e.g. tickets, pieces of cloth, etc.)

Test area load capacity (kg)

20

50

100

door

With observation window, door handle (right handle, left hinge, with key)

Window size (W x H mm)

355 x 355

Sample basket load capacity (per sample basket)

2.5kg

4.0kg

access port

φ80(ceiling)

refrigerator

Air-cooled, low-noise, 2-way refrigeration system

Water-cooled, low-noise, dual-refrigeration system, 2-way cooling water control

Refrigerant used

R449A+R23

cooler

Plate Fin Coil

High/low temperature chamber thermal fuse

1 each of high and low greenhouse

Control method

PID control

Setting/Indicator

Attached to door

Display method

7-inch LCD

Setting method

Interactive method with LCD touch panel

Upper and lower limit temperature settings

Temperature can be set for both high and low temperature rooms

Ethernet

1 pcs (100/1000MB)

USB

1 piece

Digital IO (dedicated connector)

Input 4 pcs (24V DC, 30mA) / Output 4 pcs (24V DC, ~0.5A)

Protection TYPE

IP54

Power supply (with transformer)

AC200V 3φ 50/60Hz Voltage fluctuation ±10%

Water temperature 18°C, inlet/outlet temperature difference 10 K (K/minute)

(When MIL-STD 833G is implemented)

22

44

117

Water temperature 28°C, inlet/outlet temperature difference 5 K (K/minute)

(When MIL-STD 833G is implemented)

44

87

234

Cooling water pressure

0.25 to 0.6 MPa, inlet/outlet differential pressure 0.2 MPa or higher

Cooling water temperature

+12°C to +28°C

Cooling water connection Inlet/outlet pipe diameter

Inlet Rp3/4″|Outlet Rp3/4″(female thread)

Inlet Rp 1″|Outlet Rp 1″ (female thread)

drain hose

1/2″ (ø12) hose connection

Maximum load current with transformer 50/60Hz (A)

50/55

100/110

230/254

Maximum heat removal (kW)

7.7

0.8

2

3

noise figure

58dB(A)

56dB(A)

69dB(A)

58dB(A)

72dB(A)

safeguard

Fan motor overload protection device for low-temperature chamber/high greenhouse, spindle motor overload protection device, over-temperature/over-cooling protector for sample protection, over-temperature protector for low-temperature chamber/high greenhouse for equipment protection, protective net for sample drop prevention, interlock terminal for sample energization, refrigerator pressure switch, refrigerator overheat protector, reverse phase protector for power supply, door lock mechanism, self-diagnostic function, etc.

Standard equipment (main unit/accessories)

Main unit: RJ45 connector for Ethernet, USB connector, digital I/O (4 points each, dedicated connector), over-temperature protector for sample protection, observation window (effective dimensions: W355 x H355), high temperature chamber lighting, access port (top surface: φ80, 1 piece), signal lamp, etc. Accessories: Operation manual, sample basket (1 piece), stopper for access port (1 piece), inspection report (1 set each) Access port stopper (1 pc), Inspection report (1 set each)

Supported options

Additional access ports, specimen baskets, shelves, LN2 auxiliary cooling system, additional safety devices, dehumidifier, specimen temperature sensor, temperature output mechanism, air-cooled condenser, water-cooled condenser, etc.